Section 9
Error
Code Description
E035
Incorrect Password Entered – The password that was entered into the meter was incorrect.
Verify the password and reenter it.
E036
Bad Value Entered – The value that was entered into the meter was outside of the allowable
range of values. Verify the value range and reenter the value.
E038
Calibration Log Full – The calibration log is full and must be cleared before any additional
calibration data can be saved.
E039
Calibration Log Empty – There is no calibration saved in the calibration log and a calibration
needs to be performed before any calibration data can be viewed.
E040
Data Log Full – The data log is full and must be cleared before any data points can be saved.
E041
Data Log Empty – There are no data points saved in the data log. The data log needs to be
enabled in the setup menu and a measurement needs to be logged before any data points can
be viewed.
E105
pH Auto Buffer Recognition Error – The millivolts measured for a pH buffer during calibration
are outside of the required range for automatic buffer recognition. Review the pH calibration
procedure and check that the correct buffer set was selected in the setup menu. Clean the
electrode according to the electrode user guide. Re-calibrate the electrode with fresh buffers.
E107
pH Calibration Standard Error – The millivolts measured during calibration are the same for
two buffers or the millivolts for the two buffers are too close together. There must be at least a 5
mV difference between the two buffers. Review the pH calibration procedure and verify that the
electrode was placed in the proper buffers at the appropriate times. Clean the electrode
according to the electrode user guide. Re-calibrate the electrode with fresh buffers.
E109
Bad pH Slope or Calibration Offset – Clean the electrode according to the electrode user
guide. Re-calibrate the electrode with fresh buffers.
E306
ISE Automatic Blank Error – Review the ISE calibration procedure and verify that the
electrode was placed in the proper standards at the appropriate times. Clean the electrode
according to the electrode user guide. Re-calibrate the electrode with fresh standards. If the
error persists, disable the autoblank feature in the setup menu and re-calibrate the meter
without a zero standard. See the Automatic Blank Correction Feature section for more
information.
E307
ISE Calibration Standard Error – The millivolts measured during calibration are the same for
two standards or the millivolts for the two standards are too close together. There must be at
least a 20% difference between the entered standard values. Review the ISE calibration
procedure and verify that the electrode was placed in the proper standards at the appropriate
times. Clean the electrode according to the electrode user guide. Re-calibrate the electrode
with fresh standards.
E309
Bad ISE Slope – Verify that the correct electrode ID was selected in the setup menu. Clean the
electrode per the electrode user guide. Re-calibrate the electrode with new standards.
E511
ORP Calibration Standard Error – The millivolts measured for the ORP standard during
calibration are outside of the required range for an ORP calibration. Review the ORP
calibration procedure and verify that the electrode was placed in the standard at the
appropriate time and the Thermo Scientific ORP standard was used. Clean the electrode
according to the electrode user guide. Re-calibrate the electrode with fresh ORP standard.
Thermo Scientific
Orion Dual Star pH/ISE Benchtop Meter User Manual
|
77
Section 9
|
Customer Services
Notice of Compliance
This equipment generates, uses, and can radiate radio frequency energy and if not installed
and used in accordance with the instruction manual, may cause interference to radio
communications. It has been tested and found to comply with the limits for a Class A computing
device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide
reasonable protection against such interference when operated in a commercial environment.
Operation of this equipment in a residential area is likely to cause interference, in which case
the user, at his own expense, will be required to take whatever measures may be required to
correct the interference.
“This digital apparatus does not exceed the (Class A) limits for radio noise emissions from
digital apparatus set out in the Radio Interference Regulations of the Canadian Department of
Communications.”
“Le présent appareil numérique n’ émet pas de bruits radioélectriques dépassant les limites
applicables aux appareils numériques (de la class A) prescrites dans le Règlement sur le
brouillage radioélectrique édicté par le ministère des Communications du Canada.”
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the symbol shown here.
Further information on compliance with these directives, the recyclers in your country, and
information on Thermo Scientific Orion products that may assist the detection of substances
subject to the RoHS Directive are available by contact us using the Water and Laboratory
Products (WLP) information on the back page of this manual.