Autosampler Operation 

|

 



Section 8 

  Temperature – Press the ▲ / ▼ keys to highlight Temperature and press the f2 (select) 



key. Press the ▲ / ▼ keys to highlight ATC1 or ATC2 as the temperature source of the 

ATC probe that will be used on the autosampler and press the f2 (accept) key. 

  Tray Size – Press the ▲ / ▼ keys to highlight Tray Size and press the f2 (select) key. 



Press the ▲ / ▼ keys to highlight 28 or 48 as the size of the tray that will be used on the 

autosampler and press the f2 (accept) key. 

  Rinse Beakers – Press the ▲ / ▼ keys to highlight Rinse Beakers and press the 



f2 (select) key. Press the ▲ / ▼ keys to highlight 0, 1, 2 or 3 as the number of the rinse 

beakers that will be used on the autosampler and press the f2 (accept) key. 

  Rinse Time – Press the ▲ / ▼ keys to highlight Rinse Time and press the f2 (select) key. 



Use the numeric keypad to enter the time (in seconds) that the electrodes will be rinsed in 

each rinse beaker and press the f2 (accept) key. The rinse time can be set from 1 to 59 

seconds. 

  Stirrer Speed – Press the ▲ / ▼ keys to highlight Stirrer Speed and press the f2 (select) 



key. Press the ▲ / ▼ keys to highlight Off, 1, 2, 3, 4 or 5 as the stirrer speed of the stirrer 

probe and press the f2 (accept) key. 

  Samples – Press the ▲ / ▼ keys to highlight Samples and press the f2 (select) key. 



o  To enter the number of samples that will be run on the autosampler, press the ▲ / ▼ 

keys to highlight Number of Samples and press the f2 (accept) key. Use the numeric 

keypad to enter the number of samples and press the f2 (accept) key.  

o  To set the sample ID, press the ▲ / ▼ keys to highlight Sample ID and press the 



f2 (accept) key. Press the ▲ / ▼ keys to highlight Off, Manual or Auto Incremental 

and press the f2 (accept) key. If Auto Incremental is selected, use the alphanumeric 

keypad to enter the initial sample ID and then press the f2 (accept) key. Refer to the 

meter pullout guide for the location of the letter keys on the numeric keypad. The 

meter will save the entered sample ID and add a sequential number after it. 

o  Once all of the sample settings are entered, press the f1 (back) key. 

  Once all of the autosampler settings are entered, press the f1 (back) key. 



 

 

 



Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 

69 




Section 8 

|

 



Autosampler Operation 

Running the Autosampler  

1.  Set all of the autosampler parameters in the autosampler setup menu. Refer to the 

Autosampler Setup Menu



 section for detailed instructions. 

2.  Set the meter display so only the channel or channels that will be run on the autosampler 

are displayed. Press the channel key to select a dual channel display for running both 

channels on the autosampler, a single channel display for running channel 1 on the 

autosampler or a single channel display for running channel 2 on the autosampler.  

Note: The meter will log the data points for both channels during the autosampler run if the dual 

channel display is selected. The meter will log the data points for only one channel if the single 

channel display is selected. 

3.  Ensure that all of the electrodes are connected to the meter inputs that correspond with the 

channel(s) that will be run on the autosampler. If the stirrer probe, Cat. No. 096019, will be 

used with the autosampler, connect the stirrer probe to the stirrer input on the back of the 

autosampler. 

4.  Set the method(s) for the channel(s) that will be run on the autosampler. Methods must be 

assigned to the channel(s) that will be run on the autosampler before the autosampler can 

be run. Make sure to assign the measurement mode and electrode ID for the channel(s) 

that will be run on the autosampler.  

The operator may select Current Settings as the method and the meter will use the current 

meter settings to run on the autosampler. If Current Settings is selected as the method, 

calibrations cannot be run on the autosampler and the samples will be measured using the 

last saved calibration.  

Note: A method can be selected when the autosampler setup parameters are entered, but any 

method editing must be done from the method list. Calibration values from the last calibration 

performed in the method are stored to the method, so calibrations can be performed on the 

autosampler. 

5.  To begin an autosampler run, press the auto key, verify that all of the autosampler setup 

parameters are correct and press the f3 (start) key.  

a.  During the autosampler run, press the f1 (pause) key to pause the run. Press the 

f1 (resume) key to continue the autosampler run at the point where it was paused, 

press the f2 (auto) key to access the autosampler setup parameters or press the 



f3 (end) key to cancel and end the autosampler run. 

b.  Press the f3 (end) key to abort the autosampler run. The autosampler arm will return to 

home position and the meter will return to the measurement mode. 

6.  Once the autosampler run is complete, another autosampler run can be started or the meter 

can be used as normal. The method that was selected for the previous autosampler run will 

be used for the next run, unless the operator changes the method in the autosampler setup 

menu. 

70 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 