Declaration of Conformity
Manufacturer:
Thermo Fisher Scientific Inc
Address:
Ayer Rajah Crescent
Blk 55 #04-16/24
Singapore 139949
Singapore
Hereby declares that the following product:
Orion Dual Star pH/ISE Benchtop Meter is
rated 100 to 240 VAC, 50/60 Hz, 0.5A
Equipment Class:
Measurement, control and laboratory
Orion Dual Star meters are EMC Class A
Conforms with the following directives and standards:
EN61326-1:2013
Electromagnetic Compatibility (EMC Directive)
Electrical equipment for measurement, control and laboratory use -
EMC requirements
EN61010-1:2010
Safety Standards
UL61010-1:2012
Safety requirements for electrical equipment for measurement,
CAN/CSA C22.2 No. 61010-1:2012 control and laboratory use - general requirements.
Cheow Kwang Chan
Place and Date of Issue:
QA/Regulatory Manager
14 May 2015
Singapore
Thermo Scientific
Orion Dual Star pH/ISE Benchtop Meter User Manual
|
79
Section 9
|
Customer Services
Real Time Clock Battery Maintenance
The Orion Dual Star meter uses an internal BR-1225 type, 3 V lithium coin cell battery for real
time clock operation. If the real time clock function is lost, the battery needs to be replaced.
Since the battery is inside of the meter housing, only qualified technicians should attempt to
replace the battery.
Replacing and Disposing of the Battery
1. Disconnect the universal power adapter from the meter and the wall outlet.
2. Orientate the meter so the display is face down on a soft, nonmetallic, non-scratching
surface. The bottom of the meter (with the pullout guide) should be closest to you and the
top of the meter (with the electrode and cable inputs) should be farthest from you.
3. Use a medium size Phillips head screwdriver to unscrew the four screws on the back of the
meter. Remove the back cover of the meter. Be careful not to spill any liquids, dust or other
foreign matter onto the exposed circuit board.
4. Locate the battery holder. It is a round, small, coin size battery holder that is located in the
left bottom corner of the circuit board and has a reference designator BAT41 marked in
white on the circuit board.
5. Use a small flat screwdriver to pry out the battery from the holder. Slide the new battery,
from the side, into the battery holder and be careful not to break the metal tab that holds the
battery in place.
6. Place the back cover on the meter and fasten the screws.
7. Dispose of the used battery in safe and environmentally responsible way, following all
applicable federal, state and local regulations.
Important Note: Replace the battery only with a new BR-1225 type, 3 V lithium coin cell
battery. Do not substitute any other battery type.
Battery Recommendations: Use caution, lithium can cause fire, explosion and burns. Follow
all applicable regulations for disposal of products containing lithium. Do not tamper with the
batteries or try to open them. Do not crush, recharge, incinerate or heat the batteries above 212
°F (100 °C). Do not expose the batteries to heat sources. Burning the batteries will result in
corrosive and toxic lithium hydroxide gas release.
80
|
Orion Dual Star pH/ISE Benchtop Meter User Manual
Thermo Scientific
Customer
Services
|
Section 9
Ordering Information
Cat. No.
Description
2115000
Orion Dual Star pH/ISE benchtop meter with 927007MD Orion stainless steel ATC
probe, electrode stand, USB cable, universal power supply, literature & test certificate
2115001
Orion Dual Star pH/ISE benchtop meter kit with 8102BNUWP ROSS Ultra glass-body
pH electrode, 927007MD Orion stainless steel ATC probe, 910104 Orion pH 4.01 buffer
(475 mL), 910107 Orion pH 7.00 buffer (475 mL), 910110 Orion pH 10.01 buffer (475
mL), 810001 ROSS storage solution (475 mL), electrode stand, USB cable, universal
power supply, literature & test certificate
2115101
Orion Dual Star pH/ISE benchtop meter & electrode set with 8102BNUWP ROSS Ultra
glass-body pH electrode, 927007MD Orion stainless steel ATC probe, electrode stand,
USB cable, universal power supply, literature & test certificate
2115102
Orion Dual Star pH/ISE benchtop meter & Triode set with 8157BNUMD ROSS Ultra
Triode epoxy-body pH/ATC electrode, electrode stand, USB cable, universal power
supply, literature & test certificate
2115201
Orion Dual Star pH/ISE benchtop meter pH & ammonia kit with 9512BNWP Orion
ammonia ion selective electrode, 8172BNWP ROSS Sure-Flow glass-body pH
electrode, 927007MD Orion stainless steel ATC probe, 096019 Orion Star series stirrer
probe, 951207 Orion 100 ppm ammonia standard (475 mL), 951211 Orion ammonia
ISA ionic strength adjuster (475 mL), 810199 ROSS pH buffer and solution kit,
electrode stand, USB cable, universal power supply, literature & test certificate
2115202
Orion Dual Star pH/ISE benchtop meter pH & chloride kit with 9617BNWP Orion
chloride ion selective electrode, 8172BNWP ROSS Sure-Flow glass-body pH electrode,
927007MD Orion stainless steel ATC probe, 096019 Orion Star series stirrer probe,
941708 Orion 1000 ppm chloride standard (475 mL), 940011 Orion chloride ISA ionic
strength adjuster (475 mL), 810199 ROSS pH buffer and solution kit, electrode stand,
USB cable, universal power supply, literature & test certificate
2115203
Orion Dual Star pH/ISE benchtop meter pH & fluoride kit with 9609BNWP Orion fluoride
ion selective electrode, 8172BNWP ROSS Sure-Flow glass-body pH electrode,
927007MD Orion stainless steel ATC probe, 096019 Orion Star series stirrer probe,
040908 Orion 10 ppm fluoride standard premixed with TISAB II (475 mL), 040906 Orion
1 ppm fluoride standard premixed with TISAB II (475 mL), 940909 Orion TISAB II total
ionic strength adjustment buffer (3.8 L), 810199 ROSS pH buffer and solution kit,
electrode stand, USB cable, universal power supply, literature & test certificate
2115204
Orion Dual Star pH/ISE benchtop meter pH & nitrate kit with 9707BNWP Orion nitrate
ion selective electrode, 8172BNWP ROSS Sure-Flow glass-body pH electrode,
927007MD Orion stainless steel ATC probe, 096019 Orion Star series stirrer probe,
930707 Orion 100 ppm nitrate standard (475 mL), 930711 Orion nitrate ISA ionic
strength adjuster (475 mL), 810199 ROSS pH buffer and solution kit, electrode stand,
USB cable, universal power supply, literature & test certificate
2115205
Orion Dual Star pH/ISE benchtop meter pH & sodium kit with 8611BNWP ROSS
sodium ion selective electrode, 8172BNWP ROSS Sure-Flow glass-body pH electrode,
927007MD Orion stainless steel ATC probe, 096019 Orion Star series stirrer probe,
841108 Orion 1000 ppm sodium standard (475 mL), 841109 Orion KAP sodium
standard & ISA solution (475 mL), 841111 Orion sodium ISA ionic strength adjuster
(475 mL), 841113 Orion sodium electrode conditioning solution (475 mL), 841101 Orion
sodium electrode storage solution (475 mL), 810199 ROSS pH buffer and solution kit,
electrode stand, USB cable, universal power supply, literature & test certificate