Orion Dual Star pH/ise dual Channel Benchtop Meter User Manual



Yüklə 0,73 Mb.
Pdf görüntüsü
səhifə33/33
tarix05.01.2018
ölçüsü0,73 Mb.
#19796
1   ...   25   26   27   28   29   30   31   32   33

Advanced Meter Features 

|

 



Section 10 

Automatic Blank Correction Feature 

The Orion Dual Star meter has a feature known as automatic blank correction. In this 

technique, the meter decides whether blank correction is the best measurement strategy by 

analyzing the electrode response during a multipoint calibration. A separate blank does not 

have to be run. Graphically, automatic blank correction is equivalent to drawing a smooth curve 

through the lowest three points of the multipoint calibration and extrapolating to zero 

concentration based on the assumption of Nernstian electrode behavior.  

Multipoint calibrations at the lower limit of detection are desirable when the response of an ion 

selective electrode is non-linear and cannot be characterized with a one or two point calibration. 

This is usually seen as a low electrode slope. Generally, the electrode is behaving in a 

Nernstian manner but the effect of a blank is being observed. See the figure below. 

 

 

This blank may be a true reagent blank, traces of analyte ion in the reagents or it may be the 



“mud” value of the electrode. It could also be an interference in the reagents that becomes 

apparent at low levels of analyte ion, or it could be any combination of these effects.  

The expanded version of the Nernst equation traditionally used for blank correction is as 

follows: 

E = E

o

 + S x log (C + b)  where b is the blank 



In a multi-point calibration, a set of equations is generated and the relationship between them 

evaluated. For example the equations generated in a three point calibration would be as 

follows: 

E

1



 = E

o

 + S x log (C



1

 + b) 


E

2

 = E



o

 + S x log (C

2

 + b) 


E

3

 = E



o

 + S x log (C

3

 + b) 


Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 

87 




Section 10 

|

 



Advanced Meter Features 

The meter evaluates the relationships between the three potentials E

1

, E


2

 and E


3

 and the three 

concentrations C

1

, C



2

 and C


3

. If the relationships indicate that blank correction is desired, a 

blank will automatically be calculated and the non-linearity will be corrected for in a Nernstian 

manner. If the appropriate conditions are not met, the blank is set at zero and each segment of 

the multipoint calibration is treated independently. 

When all three of the following conditions are met, blank correction is invoked. 

1.  The concentration of the first standard is zero, or the slope of the electrode between the first 

and second standards is less than the slope between the second and third. 

2.  Potential differences between points are significant. For example,  

3.  E


3

 - E


1

 > 10 mV 

4.  The blank correction algorithm converges at reasonable blank and slope values. Conditions 

in steps 1 and 2 prevent failure to converge in most situations. However, the slope will attain 

any value necessary for convergence to a calculated blank value of 3 x C

3



When blank correction is implemented, the slope value for the electrode may be outside the 

range of values normally considered acceptable during an ordinary calibration. For best results, 

calibration standards should be close in range to the expected sample concentrations and 

should bracket the expected sample concentration. 

When conditions 1, 2 and 3 are not met, the calibration data is handled by the multipoint 

calibration method described earlier. 



Note: When a calibration standard is defined as zero and the meter cannot calculate a blank, 

an error code E306 will be displayed.  

In calibrations with more than three points, a combination of methods are utilized. Automatic 

blank correction is used if the lowest three points satisfy the criteria and multipoint calibration is 

used for the other points. The slope displayed after calibration is an average of the slope 

calculated in the automatic blank correction algorithm and the slope(s) for each additional 

segment.

 

Error Code E306 

With the Autoblank Function On 

This error occurs during a three or more point calibration when one solution is defined as zero 

concentration and the actual concentration of the blank is substantially greater than the 

standard with the third highest concentration. Since the blank cannot be calculated, the meter 

cannot use the autoblank function and cannot use the zero concentration standard in the two 

point calibration segment. To clear the error, calibrate the meter without using a zero 

concentration standard. 

With the Autoblank Function Off 

This error occurs during a one or two point calibration if one standard is defined as zero 

concentration. To clear the error, calibrate the meter without using a zero concentration 

standard. 



88 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 




Advanced Meter Features 

|

 



Section 10 

Isopotential Point Feature 

To take advantage of the temperature compensation feature for concentration measurements in 

the Orion Dual Star meter, it is necessary to experimentally determine the isopotential point of 

the electrode in use. The slope of all electrodes changes with temperature. This is due in fact to 

the slope term, S, in the Nernst equation. The slope is actually 2.3 RT/nF, where R and F are 

constants, n is the charge on the measured species and T is the temperature in degrees Kelvin. 

Examples of calibration curves at varying temperatures are shown in the figure below.  

These curves intersect at the isopotential point, which is the concentration at which the potential 

of the electrode does not vary with temperature. If the isopotential point is known or can be 

measured experimentally, and if the meter has the means of adjusting the isopotential point, 

temperature compensation for an ion selective electrode is possible. Temperature 

compensation can be performed for one and two point calibrations. 

 

 



Determining an Isopotential Point 

1.  Prepare several standards with concentration ranges over the measuring range of the 

electrode. 

2.  Measure the millivolt value of the standards at room temperature, about 20 to 25 °C. 

3.  Measure the millivolt value of each standard at 75 °C. 

4.  Measure the millivolt value of each standard at 10 °C. 

5.  On semi-logarithmic graph paper, plot the concentration values on the log axis versus the 

millivolt values on the linear axis, as any calibration curve would be drawn (see the figure on 

the previous page). 

6.  The lines will intersect at the isopotential point. Read the concentration off the graph for this 

point from the log axis. This value is the isopotential point. 

7.  See the Setup Menu section for instructions on how to adjust the isopotential point on the 

Orion Dual Star meter. 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 



89 


Section 10 

|

 



Advanced Meter Features 

Automatic pH Buffer Recognition Feature 

The Orion Dual Star meter is capable of automatically recognizing pH 1.68, 4.01, 7.00, 10.01 

and 12.46 buffers or pH 1.68, 4.01, 6.86, and 9.18 buffers during a pH calibration, depending 

on the pH buffer set that is selected in the setup menu. During a calibration, the meter uses the 

selected buffer set and the raw mV reading of the pH electrode in the buffer to recognize and 

display the buffer value at the measured temperature. The raw mV reading of the pH electrode 

in the buffer must be about ± 30 mV from the theoretical mV reading of the buffer in order for 

the meter to automatically recognize the buffer. 

USA pH Buffer Set 

EURO pH Buffer Set 

Buffer 

mV Range 

 

Buffer 

mV Range 

1.68 


+285 to +345 mV 

 

1.68 



+285 to +345 mV 

4.01 


+207 to +147 mV 

 

4.01 



+207 to +147 mV 

7.00 


- 30 to + 30 mV 

 

6.86 



+38 to -22 mV 

10.01 


-207 to -147 mV 

 

9.18 



-99 to -159 mV 

12.46 


-293 to -353 mV 

 

 



 

 

Testing a pH Electrode for Automatic Buffer Recognition 

To verify that the raw mV reading of the pH electrode in use is ± 30 mV from the theoretical mV 

reading of the buffer, and therefore verify that the pH electrode in use is capable of performing 

automatic buffer recognition, perform the following procedure.  

1.  Set the measurement mode to mV in the setup menu. Prepare the pH electrode according 

to the instructions in the pH electrode user guide. Ensure that the pH electrode is connected 

to the meter input that corresponds with the channel that will be measured.  

2.  Rinse the pH electrode with distilled water, blot it dry with a lint-free tissue and place the pH 

electrode into a pH 4.01 buffer at 25 °C.  

3.  Wait for the measurement to stabilize and record the mV value of the pH 4.01 buffer when 

the measurement is stable. The read type selected in the EZ Startup menu or setup menu 

will determine how the mV measurements are displayed by the meter.  

4.  Remove the pH electrode from the pH 4.01 buffer, rinse the pH electrode with distilled 

water, blot it dry with a lint-free tissue and place it into a pH 7.00 buffer at 25 °C.  

5.  Wait for the measurement to stabilize and record the mV value of the pH 7.00 buffer when 

the measurement is stable. 

6.  The mV reading of the pH electrode in pH 4.01 buffer should be +207 to +147 mV and the 

mV reading of the pH electrode in pH 7.00 buffer should be -30 to +30 mV. If the mV 

readings are in the correct ranges, the pH electrode is capable of performing automatic 

buffer recognition. If the mV readings of the pH electrode are not in the correct ranges, the 

values of the pH buffers will need to be manually entered during a pH calibration. 



90 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 




 


 

 

 



255100-001 0615  RevC  

thermoscientific.com/water 

© 2015 Thermo Fisher Scientific Inc. All rights reserved. Microsoft and Excel are either registered trademarks or trademarks of Microsoft Corporation. All 

other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. 

North America  

Toll Free: 1-800-225-1480 

Tel: 1-978-232-6000 

Info.water@thermofisher.com 



Germany 

Tel: (49) 6184-90-6000 

info.water.uk@thermofisher.com

 

China 

Tel: (86) 21-68654588  

wai.asia@thermofisher.com

 

India 

Tel: (91) 22-4157-8800 

wai.asia@thermofisher.com

 

Singapore 

Tel: (65) 6778-6876 

wai.asia@thermofisher.com

 

Japan 

Tel: (81) 045-453-9175 

wai.asia@thermofisher.com

 

Australia 

Tel: (613) 9757-4300 

In Australia (1300) 735-295 

InfoWaterAU@thermofisher.com

 

Water and Lab Products



 

Document Outline

  • 1
  • Section 1
  • Meter Introduction
    • Meter Overview
      • Display Features
      • Data Features
      • Troubleshooting Features
      • Redesigned Orion Dual Star Meters
    • Meter Packing List
    • Intended Use
  • 2
  • Section 2
  • Meter Basics
    • Using the Universal Power Adapter
    • Attaching the Electrode Stand and Holder
      • Using the Electrode Holder
    • Powering the Meter On and Off
    • Meter Maintenance
    • Meter Components
      • Meter Connections
      • Channel 1 and Channel 2 Inputs
      • Connecting Electrodes and Cables to the Meter
      • Meter Keypad
      • Alphanumeric Keypad
      • Meter Display
  • 3
  • Section 3
  • Meter Setup
    • EZ Startup Menu
    • Setup Menu
      • Navigating the Setup Menu
      • Channel 1 and Channel 2 Setup Menus
        • Measure Mode
        • Electrode ID
        • Electrode Serial Number
        • Resolution
        • Temperature Calibration
        • Temperature Input
        • Calibration Setup
        • Read Type
        • Alarm
        • Sample ID
        • Method
      • Instrument Parameters Setup Menu
        • Export Data (PC/Printer/Log)
        • Stirrer Speed
        • Log View
        • Display
        • Calendar
        • Troubleshooting
        • Beeper Volume
        • Language
      • Autosampler Setup Menu
      • Incremental Techniques Setup Menu
      • GLP Setup Menu
      • EZ Startup Setup Menu
  • 4
  • Section 4
  • Calibration and Measurement
    • pH Calibration and Measurement
      • pH Calibration Hints
      • pH Calibration Procedure (1 to 6 points)
      • pH Measurement Procedure
    • ISE Calibration and Measurement
      • ISE Calibration Hints
      • ISE Calibration Procedure (2 to 6 points)
      • ISE Measurement Procedure
    • mV Measurement and RmV and ORP Calibration and Measurement
      • mV Measurement Procedure
      • RmV Calibration Hints
      • RmV Calibration Procedure
      • RmV Measurement Procedure
      • ORP Calibration Hints
      • ORP Calibration Procedure
      • ORP Measurement Procedure
    • Temperature Calibration and Measurement
      • Temperature Calibration Procedure
  • 5
  • Section 5
  • Methods
    • Saving the Current Meter Settings as a New Method
    • Loading an Existing Method
    • Editing or Printing an Existing Method
    • Coping an Existing Method
    • Deleting an Existing Method
  • 6
  • Section 6
  • Incremental Techniques
    • Electrode ID and Slope for Incremental Techniques
    • Single Known Addition
      • Single Known Addition Procedure
    • Double Known Addition
      • Double Known Addition Procedure
    • Single Known Subtraction
      • Single Known Subtraction Procedure
    • Double Known Subtraction
      • Double Known Subtraction Procedure
  • 7
  • Section 7
  • Data Storage and Retrieval
    • Data Storage Settings
    • Viewing the Calibration Log and Data Log
    • Interfacing a Printer
    • Interfacing a Computer
    • Meter Software Upgrade
  • 8
  • Section 8
  • Autosampler Operation
    • Interfacing the Meter and Autosampler
    • Autosampler Setup Menu
    • Running the Autosampler
      • Autosampler Error Messages
  • 9
  • Section 9
  • Customer Services
    • Troubleshooting Guide
      • Electrode Stability Test
      • Meter Self Test and Checkout
      • User Reset
      • Factory Configuration
      • About Meter
      • Electrode Condition Icon
    • Assistance
    • Warranty
    • Meter Error Codes
      • Over Range Measurements
    • Notice of Compliance
    • WEEE Compliance
    • Declaration of Conformity
    • Real Time Clock Battery Maintenance
      • Replacing and Disposing of the Battery
    • Ordering Information
    • Meter Specifications
  • 10
  • Section 10
  • Advanced Meter Features
    • Multipoint Calibration Curve Feature
    • Automatic Blank Correction Feature
      • Error Code E306
        • With the Autoblank Function On
        • With the Autoblank Function Off
    • Isopotential Point Feature
      • Determining an Isopotential Point
    • Automatic pH Buffer Recognition Feature
      • USA pH Buffer Set EURO pH Buffer Set
      • Testing a pH Electrode for Automatic Buffer Recognition

Yüklə 0,73 Mb.

Dostları ilə paylaş:
1   ...   25   26   27   28   29   30   31   32   33




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©www.genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə