Section 3
|
Meter Setup
When the meter displays Keypad Test Passed, press the f2 (next) key. The meter will
display Self Test Passed. Press the f1 (escape) key to return to the measurement mode.
•
User Reset – Press the ▲ / ▼ keys to highlight User Reset and press the f2 (accept) key.
If the user reset is performed, all of the current meter settings will be deleted and reset to
the default settings. The data log, calibration log and methods will not be deleted with the
user reset. Press the f2 (yes) key to perform the reset or press the f3 (no) key to abort the
reset. The meter will enter the EZ Startup menu and then proceed to the measurement
mode.
•
Factory Config – Press the ▲ / ▼ keys to highlight Factory Config and press the
f2 (accept) key. If the factory configuration is performed, all of the current meter settings will
be deleted and reset to the default settings and the data log, calibration log and methods
will be deleted. Press the f2 (yes) key to perform the reset or press the f3 (no) key to abort
the reset. Use the alphanumeric keypad to enter the PIN number and press the f2 (accept)
key to perform the reset. The default PIN number is 111111111. Press the f1 (measure)
key, the meter will enter the EZ Startup menu and then proceed to the measurement mode.
•
About Meter – Press the ▲ / ▼ keys to highlight About Meter and press the f2 (accept)
key. The meter serial number and software version will be displayed. Press the
f1 (back)
key to return to the troubleshooting settings and press the f1 (back) key to return to the
instrument parameter settings.
Beeper Volume
Press the ▲ / ▼ keys to highlight Beeper Volume and press the f2 (select) key. Press the
▲ / ▼ keys to highlight Off, Low or High and press the f2 (accept) key.
Language
Press the ▲ / ▼ keys to highlight Language and press the f2 (select) key. Press the ▲ / ▼
keys to highlight English, Español, Deutsch, Français or Italiano and press the f2 (accept) key.
32
|
Orion Dual Star pH/ISE Benchtop
Meter User Manual
Thermo Scientific
Meter
Setup
|
Section 3
Autosampler Setup Menu
The autosampler menu contains a list of meter parameters and settings that are used when the
meter runs the autosampler and the command to turn the autosampler on or off. For
instructions on navigating this menu, refer to Section 8 Autosampler Operation.
The incremental techniques menu allows the operator to access the single known addition,
single known subtraction, double known addition and double known subtraction meter
programs. For instructions on navigating this menu, refer to Chapter 6 Incremental Techniques.
1. In the measurement mode, press the setup key to enter the setup menu.
2. Press the ▲ / ▼ keys to highlight GLP and press the f2 (select) key.
3. The meter will prompt for the old password. Enter the current password and press the
f2 (accept) key. The default password is 111111111.
4. The meter will prompt for the new password. Enter a new password using the alphanumeric
keypad and press the f2 (accept) key.
For instructions on navigating this menu, refer to the EZ Startup Menu section.
Thermo Scientific
Orion Dual Star pH/ISE Benchtop Meter User Manual
|
33
4
Calibration and Measurement
The Orion Dual Star meter can perform a one to six point pH calibration, a two to six point ISE
(concentration) calibration, a one point ORP calibration and a one point temperature calibration.
Step-by-step instructions and prompts are displayed on the meter to assist operators with the
calibration menus. A calibration graph, which includes the electrode slope between points and
the average slope, can be viewed in the calibration mode for pH and ISE (concentration)
calibrations. The meter can save up to 10 calibrations per channel in the log and calibration
data can be exported from the meter to a printer or computer.
pH Calibration and Measurement
pH Calibration Hints
•
The measurement mode (displayed to the right of the measurement value) determines the
type of calibration that the meter will perform. Set the measurement mode to pH for the
channel that will be calibrated. Refer to the
Setup Menu
section for instructions on setting
the measurement mode.
•
Prepare and condition the pH electrode per the instructions in the electrode user guide.
•
Make sure that the pH electrode, ATC probe and reference electrode (if applicable) are
connected to the meter inputs that correspond with the channel that will be calibrated.
•
Always use fresh pH buffers for calibration. Choose buffers that are 1 to 4 pH units apart.
•
The meter will automatically recognize the US buffer set (pH 1.68, 4.01, 7.00, 10.01 and
12.46 buffers) or the EURO buffer set (pH 1.68, 4.01, 6.86 and 9.18 buffers). The default
meter setting is the US buffer set.