Section 3 

|

 



Meter Setup 

Resolution 

Press the ▲ / ▼ keys to highlight Resolution and press the f2 (select) key. Press the ▲ / ▼ 

keys to highlight the desired measurement display resolution and press the f2 (accept) key. 

This menu may not be accessible, depending on the measurement mode that was selected for 

the channel. 

Temperature Calibration 

Note: The meter ATC temperature display has a relative accuracy of ± 0.1 °C. ATC probes 

have varying temperature accuracies, usually ± 0.5 °C to ± 2 °C. Use this function only if it is 

necessary to calibrate the temperature readings. Since the temperature offset calculated during 

the calibration is applied to all future temperature measurements, recalibrate if a different ATC 

probe is used. 

Press the ▲ / ▼ keys to highlight Temperature Calibration and press the f2 (select) key. Press 

the ▲ / ▼ keys to highlight ATC1, ATC2 or ATC1 & ATC2 to select the ATC probe(s) that will 

be calibrated and press the f2 (select) key. Place the ATC probe(s) and thermometers into a 

solution with a known, stable temperature. It is recommended that two NIST traceable 

thermometers be used to measure and verify the temperature of the solution. Wait for the 

readings of the ATC probe(s) and thermometers to stabilize (about five to ten minutes); use the 

numeric keypad, decimal key and +/- key to enter the temperature value read by the 

thermometer in the Reference Temperature: ____°C field and press the f2 (accept) key. The 

meter will display the original temperature read by the ATC probe (i.e. ATC1: 24.2°C) and the 

temperature offset value that will be applied to all future temperature readings (i.e. ATC1 Offset: 

-0.8°C). Press the f3 (log/print) key to save and end the calibration, export the data to the 

calibration log, export the data to a printer or computer if one is connected to the meter and 

enabled in the setup menu, and proceed to the measurement mode or press the f1 (escape) 

key to return to the measurement mode without saving the calibration. 

Temperature Input 

Press the ▲ / ▼ keys to highlight Temperature Input and press the f2 (select) key. Select the 

source of the temperature measurement and the display units of the temperature measurement.  

  ATC1, ATC2 or Manual – To select the temperature source, press the ▲ / ▼ keys to 



highlight ATC1, ATC2 or Manual and press the f2 (accept) key. Channel 1 and 2 are able 

to receive temperature readings from an ATC probe connected to the channel 1 input 

(ATC1) or the channel 2 input (ATC2). If Manual is selected, enter the temperature using 

the numeric keypad, decimal key and +/- key and press the f2 (accept) key.  

  Temperature Unit – To select the temperature units as Celsius or Fahrenheit, press the 



▲ / ▼ keys to highlight Temperature Unit, press the f2 (accept) key, press the ▲ / ▼ 

keys to highlight Celsius or Fahrenheit and press the f2 (accept) key. 

Once all of the temperature settings are entered, press the f1 (back) key. 

24 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 




Meter Setup 

|

 



Section 3 

Calibration Setup 

Press the ▲ / ▼ keys to highlight Calibration Setup and press the f2 (select) key. View and 

change the pH or ISE calibration settings, depending on the measurement mode that was 

selected for the channel. 

  pH Calibration Settings – Entering the Calibration Setup menu will access the Buffer Set 



and ISO settings.  

  Buffer Set – Press the ▲ / ▼ keys to highlight Buffer Set and press the f2 (select) key. 



Press the ▲ / ▼ keys to select US or EURO as the calibration buffer set and press the 

f2 (accept) key. The meter will automatically recognize pH 1.68, 4.01, 7.00, 10.01 and 

12.46 buffers if US is selected and pH 1.68, 4.01, 6.86 and 9.18 buffers if EURO is 

selected.  

  ISO – Press the ▲ / ▼ keys to highlight ISO and press the f2 (select) key. To enter the pH 



isopotential point, use the numeric keypad and decimal key to enter the value and press 

the f2 (accept) key. 

Once all of the pH calibration settings are entered, press the f1 (back) key. 

  ISE calibration settings – Entering the Calibration Setup menu will access the Autoblank, 



Low Level Stability and Isopotential settings. 

o  Autoblank – Press the ▲ / ▼ keys to highlight Autoblank and press the f2 (select) 

key. Press the ▲ / ▼ keys to highlight Off or On and press the f2 (accept) key.  

o  Low Level Stability – Press the ▲ / ▼ keys to highlight Low Level Stability and press 

the f2 (select) key. Press the ▲ / ▼ keys to highlight Off or On and press the 

f2 (accept) key.  

o  Isopotential – Press the ▲ / ▼ keys to highlight Isopotential and press the f2 (select) 

key.  

o  To turn off or on temperature compensation for ISE measurements, press the ▲ / ▼ 



keys to highlight ISO On/Off and press the f2 (select) key. Press the ▲ / ▼ keys to 

highlight Off or On and press the f2 (accept) key. 

o  To enter the ISE isopotential point, press the ▲ / ▼ keys to highlight ISO Value and 

press the f2 (select) key. Use the numeric keypad and decimal key to enter the value 

and press the f2 (accept) key. 

o  Once all of the ISE isopotential settings are entered, press the f1 (back) key. 

Once all of the ISE calibration settings are entered, press the f1 (back) key. 

Read Type 

Press the ▲ / ▼ keys to highlight Read Type and press the f2 (select) key. Press the ▲ / ▼ 

keys to highlight Auto-Read, On Ready, At Time Intervals, Continuous, Value Change In 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 



25 