Orion Dual Star pH/ise dual Channel Benchtop Meter User Manual



Yüklə 0,73 Mb.
Pdf görüntüsü
səhifə16/33
tarix05.01.2018
ölçüsü0,73 Mb.
#19796
1   ...   12   13   14   15   16   17   18   19   ...   33

Calibration and Measurement 

|

 



Section 4 

ISE Measurement Procedure 

1.  Rinse the ISE electrode (ATC probe, stirrer probe and reference electrode, if applicable) 

with distilled water and place the electrode(s) into the sample. 

2.  If the stirrer probe is in use, press the stirrer key to turn on the stirrer probe. Press the 

▲ / ▼ keys to adjust the speed of the stirrer probe. 

3.  Wait for the measurement to stabilize and record the concentration and temperature of the 

sample when the meter indicates that the measurement is stable. The read type selected in 

the EZ Startup menu or setup menu will determine how the ISE measurements are 

displayed by the meter. The available read types are Auto-Read, on ready, at time intervals, 

continuous, value change in measurement and timed reading. To review or change the 

available read type, refer to the 

Setup Menu

 section.  

4.  If the stirrer probe is in use, press the stirrer key to turn off the stirrer probe. 

5.  Remove the electrode(s) from the sample, rinse with distilled water, blot dry, place the 

electrode(s) into the next sample. 

6.  Repeat steps 2 through 5 for all of the samples. 

7.  When all of the samples have been measured, store the electrode(s) according to the 

instructions in the electrode user guides. 

 

 



 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 



39 


Section 4 

|

 



Calibration and Measurement 

mV Measurement and RmV and ORP 

Calibration and Measurement 

mV Measurement Procedure 

Measure the raw millivolt (mV) values of an ORP, ISE or pH electrode in the mV mode. 

Oxidation-reduction potential (ORP) measurements are commonly reported as raw millivolts. 

The mV readings of an ISE electrode in standards can be used to draw a calibration curve on 

semi-logarithmic graph paper, which can be helpful for low level ISE measurements in the non-

linear range of the electrode. The millivolt readings of a pH electrode can be used to 

troubleshoot the electrode and verify it is reading the correct mV range of common pH buffers. 

1.  Set the measurement mode to mV in the setup menu. Prepare the electrode according to 

the instructions in the electrode user guide. Ensure that the electrode, ATC probe and 

reference electrode (if applicable) are connected to the meter inputs that correspond with 

the channel that will be measured.  

2.  Rinse the electrode (ATC probe, stirrer probe and reference electrode, if applicable) with 

distilled water, blot dry with a lint-free tissue and place the electrode(s) into the sample. 

3.  If the stirrer probe is in use, press the stirrer key to turn on the stirrer probe. 

4.  Wait for the measurement to stabilize and record the mV and temperature of the sample 

when the meter indicates that the measurement is stable. The read type selected in the EZ 

Startup menu or setup menu will determine how the mV measurements are displayed by the 

meter. The available read types are Auto-Read, on ready, at time intervals, continuous, 

value change in measurement and timed reading. To review or change the available read 

type, refer to the 

Setup Menu

 section.  

5.  If the stirrer probe is in use, press the stirrer key to turn off the stirrer probe. 

6.  Remove the electrode(s) from the sample. 

7.  Repeat steps 2 through 6 for all of the samples. 

8.  When all of the samples have been measured, store the electrode(s) according to the 

instructions in the electrode user guides. 

 

 



 

40 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 




Calibration and Measurement 

|

 



Section 4 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 



41 

RmV Calibration Hints 

  The measurement mode (displayed to the right of the measurement value) determines the 



type of calibration that the meter will perform. Set the measurement mode to RmV for the 

channel that will be calibrated. Refer to the 

Setup Menu

 section for instructions on setting 

the measurement mode. 

  Prepare and condition the electrode per the instructions in the electrode user guide.  



  Ensure that the electrode, ATC probe and reference electrode (if applicable) are connected 

to the meter inputs that correspond with the channel that will be calibrated.  

  Always use a freshly prepared standard for calibration. Allow the standard and samples to 



reach the same temperature for precise measurements. 

 

RmV Calibration Procedure 



Note: In most calibration screens, press the f1 (escape) key to return to the measurement 

mode without saving the calibration. 

1.  In the measurement mode, press the cal key. 

2.  Dual channel display only: Press the ▲ / ▼ keys to select the channel to be calibrated and 

press the f2 (accept) key.  

3.  Rinse the electrode (ATC probe, stirrer probe and reference electrode, if applicable) with 

distilled water, blot dry with a lint-free tissue and place the electrode(s) into the standard. 

4.  When the electrode(s) and standard are ready, press the f3 (start) key. 

5.  Wait for the value to stabilize and the meter to display ready next to the raw mV value. The 

raw mV value is the actual mV reading of the electrode. To enter a relative mV value, use 

the numeric keypad, decimal key and +/- key to enter the mV value of the standard. Press 

the f3 (clear) key to delete any digits. Once the value is correct, press the f2 (accept) key.  

6.  The meter will display the mV offset. Perform one of the following actions: 

Important Note: The f2 (log/print) key must be pressed to save the 

calibration. Only pressing the f3 (cal done) key without pressing the 



f2 (log/print) key will not save the calibration. 

a.  Press the f2 (log/print) key to save and end the calibration, export the calibration data to 

the calibration log and export the calibration data to a printer or computer, if a printer or 

computer is connected to the meter and enabled in the setup menu. The meter will 

automatically proceed to the measurement mode. 

b.  Press the f1 (escape) key to return to the measurement mode without saving the 

calibration.  

7.  Proceed to the RmV Measurement Procedure section. 




Yüklə 0,73 Mb.

Dostları ilə paylaş:
1   ...   12   13   14   15   16   17   18   19   ...   33




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©www.genderi.org 2024
rəhbərliyinə müraciət

    Ana səhifə