Section 3
|
Meter Setup
Instrument Parameters Setup Menu
In the measurement mode, press the setup key to enter the setup menu. Press the ▲ / ▼ keys
to highlight Instrument Parameters and press the f2 (select) key. Perform the following actions,
as required.
Export Data (PC/Printer/Log)
Press the ▲ / ▼ keys to highlight Export Data (PC/Printer/Log) and press the f2 (select) key.
View and change the data exporting settings, if required.
•
Export Trigger – Press the ▲ / ▼ keys to highlight Export Trigger and press the f2 (select)
key. Press the ▲ / ▼ keys to highlight Channel 1, Channel 2 or Channel 1 & 2 and press
the f2 (accept) key. The read type of the selected channel(s) will determine when the meter
sends the measurement data to the data log, if the data log is enabled in the setup menu,
and to a printer or computer, if a printer or computer is connected to the meter and enabled
in the setup menu.
•
PC – Press the ▲ / ▼ keys to highlight PC and press the f2 (select) key. View and change
the PC settings, if required.
o PC On/Off – Press the ▲ / ▼ keys to highlight PC On/Off and press the f2 (select)
key. To enable communication to a computer, press the ▲ / ▼ keys to highlight On
and press the f2 (accept) key. The meter will send the data to a computer in a comma
delimited format. To disable communication to a computer, press the ▲ / ▼ keys to
highlight Off and press the f2 (select) key.
o Baud Rate – Press the ▲ / ▼ keys to highlight Baud Rate and press the f2 (select)
key. Press the ▲ / ▼ keys to highlight a baud rate of 1200, 2400, 4800, 9600, 19200
or 38400 and press the f2 (accept) key.
o Once all of the PC settings are entered, press the f1 (back) key.
•
Printer – Press the ▲ / ▼ keys to highlight Printer and press the f2 (select) key. View and
change the printer settings, if required.
o Printer On/Off – Press the ▲ / ▼ keys to highlight Printer On/Off and press the
f2 (select) key. To enable communication to a printer, press the ▲ / ▼ keys to
28
|
Orion Dual Star pH/ISE Benchtop
Meter User Manual
Thermo Scientific
Meter
Setup
|
Section 3
highlight On and press the f2 (accept) key. To disable communication to a printer,
press the ▲ / ▼ keys to highlight Off and press the f2 (accept) key.
o Baud Rate – Press the ▲ / ▼ keys to highlight Baud Rate and press the f2 (select)
key. Press the ▲ / ▼ keys to highlight a baud rate of 1200, 2400, 4800, 9600, 19200
or 38400 and press the f2 (accept) key.
o Once all of the printer settings are entered, press the f1 (back) key.
•
Data Log – Press the ▲ / ▼ keys to highlight Data Log and press the f2 (select) key. View
and change
the data log settings, if required.
o Data Log On/Off – Press the ▲ / ▼ keys to highlight Data Log On/Off and press the
f2 (select) key. To enable the data log, press the ▲ / ▼ keys to highlight On and
press the f2 (accept) key. To disable the data log, press the ▲ / ▼ keys to highlight
Off and press the f2 (accept) key.
o Ring Buffer – Press the ▲ / ▼ keys to highlight Ring Buffer and press the f2 (select)
key. Press the ▲ / ▼ keys to highlight Off or On and press the f2 (accept) key. If Off
is selected, an error message will be displayed when all 1000 data log points are filled
and the data log must be emptied to clear the error message. If On is selected, the
oldest data log point will be overwritten with the new data log point when all 1000 data
log points are filled.
o Once all of the data log settings are entered, press the f1 (back) key.
•
Once all of the exporting settings are entered, press the f1 (back) key.
Stirrer Speed
Press the ▲ / ▼ keys to highlight Stirrer Speed and press the f2 (select) key. Press the ▲ / ▼
keys to set the default stirrer speed to 1 (slowest speed), 2, 3, 4 or 5 (fastest speed) and press
the f2 (accept) key. The default stirrer speed will be used whenever the stirrer key is pressed in
the measurement mode.
Log View
Press the ▲ / ▼ keys to highlight Log View and press the f2 (select) key. Select Data Log or
Calibration Log to view, export or clear the selected log.
•
Data Log – Press the ▲ / ▼ keys to highlight Data Log and press the f2 (accept) key. The
meter will display a list of the data points. The list shows the sequential number of the data
point (D0001), the date that the data point was saved (07/05/2008) and the time that the
data point was saved (14:05). To view the measurement information for an individual data
point, press the ▲ / ▼ keys to highlight the data point and press the f2 (select) key. Press
the f2 (print) key to print the data point or press the f1 (back) key to return to the list of data
points. To export the data log to a computer or printer or to clear the data log, press the
f3 (options) key and select Log Export or Log Clear.
Thermo Scientific
Orion Dual Star pH/ISE Benchtop Meter User Manual
|