Section 3 

|

 



Meter Setup 

Measurement or Timed Reading as the read type that will be used to display the measurements 

for the selected channel and press the f2 (select) key. 

  Auto-Read – The meter will display the measurement as it stabilizes and lock and hold the 



measurement when it is stable. Once a measurement is locked, press the measure key to 

take a new measurement. Once the measurement is locked, the meter will automatically 

export the measurement to the data log, if the data log is enabled in the setup menu, and 

to a printer or computer, if a printer or computer is connected to the meter and enabled in 

the setup menu. 

  On Ready – The meter will show stabilizing on the display until the measurement is stable. 



When the measurement is stable, the meter will show ready on the display. The display will 

automatically update if the measurement changes and the measurement will not lock. 

Once ready is shown on the display, the meter will automatically export the measurement 

to the data log, if the data log is enabled in the setup menu, and to a printer or computer, if 

a printer or computer is connected to the meter and enabled in the setup menu. 

  At Time Intervals – The meter will read and display a measurement at the set time interval 



that is programmed by the operator. At the set time interval, the meter will automatically 

export the measurement to the data log, if the data log is enabled in the setup menu, and 

to a printer or computer, if a printer or computer is connected to the meter and enabled in 

the setup menu. If At Time Intervals is selected, enter the time interval value in a hours : 

minutes : seconds format using the numeric keypad and press the f2 (accept) key. 

  Continuous – The meter will continuously measure and update the display. No indicator will 



be shown on the display. This read type is useful when performing an experiment that 

requires continuous measurements to be taken, regardless of the measurement stability. 

Press the log/print key to export the measurement to the data log, if the data log is enabled 

in the setup menu, and to a printer or computer, if a printer or computer is connected to the 

meter and enabled in the setup menu. 

  Value Change In Measurement – The meter will display the measurement when the 



reading reaches or exceeds the set high or low value that is programmed by the operator. 

When the reading reaches or exceeds the set high or low value, the meter will 

automatically export the measurement to the data log, if the data log is enabled in the 

setup menu, and to a printer or computer, if a printer or computer is connected to the meter 

and enabled in the setup menu. If Value Change In Measurement is selected, press the 

▲ / ▼ keys to highlight Min Value, enter the minimum measurement value using the 

numeric keypad and the decimal key, press the ▲ / ▼ keys to highlight Max Value, enter 

the maximum measurement value using the numeric keypad and the decimal key and 

press the f3 (next) key. 

  Timed Reading – The meter will display a measurement after a set time delay and lock and 



hold the measurement after the time delay is reached. To start another time delay cycle, 

press the measure key. Once the measurement is locked, the meter will automatically 

export the measurement to the data log, if the data log is enabled in the setup menu, and 

26 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 




Meter Setup 

|

 



Section 3 

to a printer or computer, if a printer or computer is connected to the meter and enabled in 

the setup menu. If Timed Reading is selected, enter the time delay value in a minutes : 

seconds format using the numeric keypad and press the f2 (accept) key. 



Alarm 

Press the ▲ / ▼ keys to highlight Alarm and press the f2 (select) key. View and change the 

alarm settings, if required.  

  Alarm – Press the ▲ / ▼ keys to highlight Alarm and press the f2 (select) key. Press the 



▲ / ▼ keys to highlight Off, Low Limit, High Limit or Both Low/High and press the 

f2 (select) key.  

  Parameter – Press the ▲ / ▼ keys to highlight Parameter and press the f2 (select) key. 



Press the ▲ / ▼ keys to highlight pH, ISE-ppm, ISE-mg/L, ISE-%, ISE-M, ISE-unitless, 

mV, RmV, ORP or °C (°F will be shown instead of °C if it is the selected temperature unit) 

and press the f2 (accept) key.  

  Low Limit – Press the ▲ / ▼ keys to highlight Low Limit and press the f2 (select) key. To 



enter a low limit alarm value, use the numeric keypad and decimal key to enter the low 

value and press the f2 (accept) key. 

  High Limit – Press the ▲ / ▼ keys to highlight High Limit and press the f2 (select) key. To 



enter a high limit alarm value, use the numeric keypad and decimal key to enter the high 

value and press the f2 (accept) key. 



Note: The read type must be set to continuous to use the alarm settings. 

Sample ID 

Press the ▲ / ▼ keys to highlight Sample ID and press the f2 (select) key. Press the ▲ / ▼ 

keys to highlight Off, Manual or Auto Incremental and press the f2 (accept) key. If Auto 

Incremental is selected, use the alphanumeric keypad to enter the initial sample ID and press 

the f2 (accept) key. The meter will save the entered sample ID and add a sequential number 

after it. 



Method 

Press the ▲ / ▼ keys to highlight Method and press the f2 (select) key. Press the ▲ / ▼ keys 

to highlight the desired method number, press the f2 (load) key and press the f2 (yes) key. To 

view a method, edit a method, copy an existing method to a new method, delete a method, print 

a method or save the current meter settings as a new method, refer to Chapter 5 Methods

 

 



 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 



27 