Calibration and Measurement 

|

 



Section 4 

ISE Calibration and Measurement 



ISE Calibration Hints 

  The measurement mode (displayed to the right of the measurement value) determines the 



type of calibration that the meter will perform. Set the measurement mode to ISE for the 

channel that will be calibrated and set the displayed ISE units as ppm, mg/L, %, M, or 

unitless. Refer to the 

Setup Menu

 section for instructions on setting the measurement 

mode and displayed units. 

  Prepare and condition the ISE electrode per the instructions in the electrode user guide.  



  Make sure that the ISE electrode, ATC probe and reference electrode (if applicable) are 

connected to the meter inputs that correspond with the channel that will be calibrated.  

  Prepare at least two standards that bracket the expected sample range and differ in 



concentration by a factor of ten.  

  Always use freshly prepared standards for calibration. 



  Allow all standards and samples to reach the same temperature for precise measurements. 

  Stir all standards and samples at a uniform, moderate rate. Place a piece of insulating 



material, such as Styrofoam or cardboard, between the magnetic stir plate and beaker to 

prevent measurement errors from the transfer of heat to the sample.  

  Always rinse the electrode with distilled water between measurements and blot or shake 



the electrode to remove the water and prevent sample carryover. Do not wipe or rub the 

electrode sensing element. 

 

ISE Calibration Procedure (2 to 6 points) 

Note: In most calibration screens, press the f1 (escape) key to return to the measurement 

mode without saving the calibration. 

1.  In the measurement mode, press the cal key. 

2.  Dual channel display only: Press the ▲ / ▼ keys to select the channel to be calibrated and 

press the f2 (accept) key.  

3.  Rinse the ISE electrode (ATC probe, stirrer probe and reference electrode, if applicable) 

with distilled water, blot dry with a lint-free tissue and place the electrode(s) into the 

standard.  

4.  When the electrode and standard are ready, press the f3 (start) key. 

5.  Wait for the concentration value to stop flashing. Use the numeric keypad and decimal key 

to enter the concentration of the standard at the measured temperature. Press the f3 (clear) 

key to delete any digits. Once the concentration value is correct, press the f2 (accept) key. 

Thermo Scientific 

Orion Dual Star pH/ISE Benchtop Meter User Manual 

|

 

37 




Section 4 

|

 



Calibration and Measurement 

38 

|

 



Orion Dual Star pH/ISE Benchtop Meter User Manual 

Thermo Scientific 

6.  Press the f2 (next) key to proceed to the next standard and repeat steps 3 through 5 or 

press the f3 (cal done) key to end the calibration.  

7.  Once the f3 (cal done) key is pressed, the meter will display a calibration summary including 

the average slope. Perform one of the following actions: 



Important Note: The f2 (log/print) key must be pressed to save the 

calibration. Only pressing the f3 (cal done) key without pressing the 



f2 (log/print) key will not save the calibration. 

a.  Press the f2 (log/print) key to save and end the calibration, export the calibration data to 

the calibration log and export the calibration data to a printer or computer, if a printer or 

computer is connected to the meter and enabled in the setup menu. The meter will 

automatically proceed to the measurement mode. 

b.  Press the f3 (graph) key to view a point-by-point graph of the calibration. After viewing 

the graph, press the f1 (back) key to return to the calibration summary, press the 

f3 (log/print) key to save and end the calibration, export the calibration data to the 

calibration log, export the calibration data to a printer or computer (if applicable) and 

automatically proceed to the measurement mode or press the f2 (escape) key to return 

to the measurement mode without saving the calibration.  

c.  Press the f1 (cal edit) key to edit the calibration points. Press the ▲ / ▼ keys to 

highlight the calibration point to be edited and press the f2 (select) key. Press the 

▲ / ▼ keys to highlight Remeasure, Edit or Delete and press the f2 (select) key.  

i.  If Remeasure is selected, repeat steps 3 through 5. Press the ▲ / ▼ keys to 

highlight another calibration point to be edited and press the f2 (select) key or 

press the f1 (back) key to return to the calibration summary. 

i.  If Edit is selected, use the numeric keypad and decimal key to enter a new 

concentration value and press the f2 (accept) key. Press the ▲ / ▼ keys to 

highlight another calibration point to be edited and press the f2 (select) key or 

press the f1 (back) key to return to the calibration summary. 

ii.  If Delete is selected, press the f2 (yes) key to delete the selected calibration point. 

If deleting the calibration point changes the calibration to a one point calibration, 

enter the electrode slope using the numeric keypad and decimal key and press the 

f2 (accept) key. Press the ▲ / ▼ keys to highlight another calibration point to be 

edited and press the f2 (select) key or press the f1 (back) key to return to the 

calibration summary. 

8.  Proceed to the ISE Measurement Procedure section. 

 

 