Meter
Setup
|
Section 3
connected to the meter and enabled in the setup menu. If Timed Reading is
selected, enter the time delay value in a minutes : seconds format using the
numeric keypad and press the f2 (accept) key.
4. Repeat step 3 for channel 2.
5. Press the ▲ / ▼ keys to highlight Printer, Computer or None as where the meter will
export the measurement and calibration data and press the f2 (accept) key. If Printer or
Computer is selected, the meter will prompt the operator to set the baud rate. Press the
▲ / ▼ keys to highlight 1200, 2400, 4800, 9600, 19200 or 38400 as the baud rate and
press the f2 (accept) key.
6. Press the f2 (yes) key to enable the exporting of the measurement data to the data log or
press the f3 (no) key to disable the exporting of the measurement data to the data log.
7. Press the f2 (measure) key to save the EZ Startup menu settings and proceed to the
measurement mode or press the f1 (back) key to edit any of the EZ Startup settings.
Thermo Scientific
Orion Dual Star pH/ISE Benchtop Meter User Manual
|
21
Section 3
|
Meter Setup
Setup Menu
The setup menu can be accessed in the measurement mode by pressing the
setup key or by
pressing the f3 (setup) key in the dual channel display mode.
The setup menu contains seven submenus: Channel 1, Channel 2, Instrument Parameters,
Autosampler, Incremental Techniques, GLP and EZ Startup. Settings for each menu can be
selected and saved in the setup menu.
1. In the setup menu, press the ▲ / ▼ keys to highlight the desired submenu and press the
f2 (select) key to view the menu options.
2. Press the ▲ / ▼ keys to highlight the desired menu option and press the f2 (select) key to
view the current setting and change the selected setting, if required.
3. If a new setting must be selected from a list of options, press the ▲ / ▼ keys to highlight
the desired setting and press the f2 (accept) key. If a new numeric value must be entered,
use the numeric keypad, decimal key and +/- key as required and press the f2 (accept) key.
4. Press the f1 (back) key to return to the list of setup submenus.
5. Press the f1 (back) key to return to the measurement mode.
22
|
Orion Dual Star pH/ISE Benchtop Meter User Manual
Thermo Scientific
Meter
Setup
|
Section 3
Channel 1 and Channel 2 Setup Menus
Note: Channel 1 and Channel 2 have the same options. However, the options must be set
separately for each channel.
In the measurement mode, press the setup key to enter the setup menu. Press the ▲ / ▼ keys
to highlight Channel 1 or Channel 2 and press the f2 (select) key. Perform the following actions,
as required.
Measure Mode
Press the ▲ / ▼ keys to highlight Measure Mode and press the f2 (select) key. Press the
▲ / ▼ keys to highlight pH, ISE, mV, RmV or ORP as the measurement mode and press the
f2 (accept) key. If ISE is selected, press the ▲ / ▼ keys to highlight ppm (parts per million),
mg/L (milligrams per liter, % (percent), M (moles per liter) or unitless (no units) as the ISE
measurement unit that will be displayed and press the f2 (accept) key.
Electrode ID
Press the ▲ / ▼ keys to highlight Electrode ID and press the f2 (select) key. Press the ▲ / ▼
keys to highlight the type of electrode and press the f2 (accept) key. The type of electrode can
be set as pH, ORP, fluoride (F-), ammonia (NH
3
), ammonium (NH
4
+), nitrate (NO
3
-), nitrite
(NO
2
-), sulfide (S
-2
), chloride (Cl-), chlorine (Cl
2
), bromide (Br-), iodide (I-), cyanide (CN-),
sodium (Na+), potassium (K+), calcium (Ca
+2
), silver (Ag+), copper (Cu
+2
), lead (Pb
+2
),
cadmium (Cd
+2
), perchlorate (ClO
4
-), fluoroborate (BF
4
-), thiocyanate (SCN-), nitrogen oxide
(NO
X
), carbon dioxide (CO
2
), oxygen (O
2
), monovalent cation (X+), divalent cation (X
+2
),
monovalent anion (X-) or divalent anion (X
-2
). Some or all of the electrode ID options may not
be accessible, depending on the selected measurement mode for the channel.
Electrode Serial Number
Press the ▲ / ▼ keys to highlight Electrode Serial Number and press the f2 (select) key. Use
the alphanumeric keys to enter a unique number and letter combination for the electrode. The
meter will automatically advance to the next digit when no keys are pressed for three seconds.
Press the f3 (clear) key to delete any of the numbers or letters. Once the serial number is
entered, press the f2 (accept) key. Refer to the meter pullout guide for the location of the letter
keys on the numeric keypad.
Thermo Scientific
Orion Dual Star pH/ISE Benchtop Meter User Manual
|
23