Water Jacketed CO2 Incubator        2-7
Thermo Scientific
Section 2
Calibration
Scroll for Parameters
Silence
Mode
Mode
Mode
Mode
Mode
CALIBRATE
TEMPCAL       XX.X
IR CAL    XX.X
O2 CAL       XX.X
RH CAL         XX
36.9
4.9
Enter
Enter
Enter
Enter
Enter
Numbers increase
Numbers increase
Numbers increase
Numbers increase
Numbers increase
Press MODE to move
to SYS CONFIG mode
Numbers decrease
Numbers decrease
Numbers decrease
Numbers decrease
Numbers decrease
Run
Set
Cal
Config
Heat
Temp
Inject
% CO
2
Scroll for Parameters
Scroll for Parameters
Scroll for Parameters
Scroll for Parameters
Scroll for Parameters
Mode
CO2 CAL    XX.X
Enter
Press         to return
to previous parameter
Press         to return
to previous parameter
Press         to return
to previous parameter
Percent O
2
Percent RH
Zero CO
2
Operating
Temperature
Press MODE to light CAL
To Calibrate:
Span Co
2
(when O option is in use)
2
(when IR sensor is in use)
(when Thermal Conductivity
sensor is in use)
(when RH option is in use)
Calibrate Mode
Chart 2-1
Press Enter to
save the setting
Press Enter to
save the setting
Press Enter to
save the setting
Press Enter to
save the setting
Press Enter to
save the setting
Press         to return
to previous parameter
Press         to return
to previous parameter


 


New HEPA Filter
Water Jacketed CO
2
Incubator        3-1
Thermo Scientific
Section 3  
Configuration
There are many features available in Configuration Mode that allow
custom setup of the incubator. These features are listed and described
below. All features may not be necessary in all applications, but are
available if needed. To enter Configuration Mode, press the Mode key
until the Config indicator lights. Press the right or left arrow until the
appropriate parameter appears in the message center. See Chart 3-1 for
more detail.
The audible alarm can be turned on or off.  The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Audible XXX is displayed in message center.
3. Press the up/down arrow to toggle Audible ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run Mode or right/left to go to
next/previous parameter.
When the REPLACE HEPA reminder is displayed and the visual alarm
flashes, the specified time has elapsed and the HEPA filter should be
replaced. To clear the display and reset the timer after replacing the HEPA
filter with a new one, follow the steps below.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until NEW HEPA  is displayed in message center.
3. Press Enter to restart the timer and clear the REPLACE HEPA alarm.
4. Press the Mode key to return to Run Mode.
Turning the Audible
Alarm ON/OFF


Setting an Access
Code
3-2
Water Jacketed CO
2
Incubator
Thermo Scientific
Section 3
Configuration
A HEPA filter replacement timer can be set for a specific amount of time,
from 1 to 12 months of actual unit running time. Time will not accrue
when the unit is turned off. The default time is 6 months. When the
allotted time has run out, REPLACE HEPA appears in the display and the
visual alarm flashes. To set the reminder, use the following procedure.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until REPL HEPA XX is displayed.
3. Press the up/down arrow to choose the number of months desired.
4. Press Enter to save the number.
5. Press the Mode key to return to Run Mode or right/left to go to
next/previous parameter.
Note After the reminder has been set, check the allotted time remaining by
going to Config Mode, then pressing the right arrow until NEW HEPA
XXX displays. This number is the remaining days before the filter
replacement time specified runs out. For example, if 12 months was chosen
in the REPL HEPA XX message screen, the NEW HEPA number would
be 365 days. 

A three-digit Access Code can be entered to avoid unauthorized personnel
from changing the setpoints, calibration, or configuration.  A setting of
000 will bypass the access code.  The factory setting is 000.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Acc Code XXX  is displayed in the message
center.
3. Press the up/down arrow to change the access code.
4. Press Enter to save the access code.
5. Press the Mode key to return to Run Mode or right/left to go to
next/previous parameter.
Setting the REPLACE
HEPA Filter Reminder 


Setting High Temp
Alarm Limit
Water Jacketed CO
2
Incubator        3-3
Thermo Scientific
Section 3
Configuration
The low temp alarm limit (tracking alarm) is the deviation from the
temperature setpoint, which will cause a low temp alarm. The low temp
alarm is variable from 0.5° below setpoint to 5.0° below setpoint.  The
factory setting is 1.0° below setpoint. A minus sign (-) in the display
indicates that the alarm setting is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Temp LO -X.X  is displayed in the message
center.
3. Press the up/down arrow to change the low temp alarm limit.
4. Press Enter to save the low temp alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
The high temp alarm limit (tracking alarm) is the deviation from the
temperature setpoint that will cause a high temp alarm. It should be noted
that this varies from the Overtemp setpoint, in that the Overtemp setpoint
configures an independent system that monitors temperature and shuts
down the system heaters if necessary.  The high temp alarm limit is simply
set to enable an audible and visual alarm that notifies the user of a
problem.  The high temp alarm is variable from 0.5° above setpoint to 5.0°
above setpoint.  The factory setting is 1.0° above setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Temp Hi X.X  is displayed in the message
center.
3. Press the up/down arrow to change the high temp alarm limit.
4. Press Enter to save the high temp alarm limit.
5. Press the Mode key to return to Run Mode or right/left to go to
next/previous parameter.
Setting Low Temp
Alarm Limit