7-2
Water Jacketed CO
2
Incubator
Thermo Scientific
Section 7
Specifications
Unit Heat Load
115V/230V
344 BTUH (100 Watt)
Shelves
Dimensions
18.5” x 18.5” (47.0cm x 47.0cm)
Construction
Stainless steel, perforated, electropolished
Surface area
2.4 sq. ft. (0.22 sq. m) per shelf
Max. per Chamber 38.4 sq.ft. 3.6 sq. m)
Loading
35 lbs (16kg) slide in and out,
50 lbs (23kg) stationary
Standard
4
Maximum
16
Construction
Water Jacket Volume
11.7 gal. (43.5 liters)
Interior Volume
6.5 cu. ft. (184.1 liters)
Interior
Type 304, mirror finish, stainless steel
Exterior
18 gauge, cold rolled steel, powder coated
Outer Door Gasket Four-sided, molded magnetic vinyl
Inner Door Gasket
Removable, cleanable, feather-edged, silicone
Electrical
115V Models
90-125VAC, 50/60 Hz, 1 PH, 3.6 FLA
230V Models
180-250VAC, 50/60 Hz, I PH, 2.0 FLA
Circuit Breaker/Power Switch
6 Amp/2 Pole
Convenience Receptacle
75 Watts max. (one per chamber)
Alarm Contacts
Power interruption, deviation of temp., CO
2

O
2
, and RH, customer connections through jack 
on back of unit. 30V, 1A max.
Optional Data Outputs
RS-485, 0-1V, 0-5V, 4-20mA
Installation Category
Overvoltage Category II
Pollution Degree 2
Maximum Leakage Current With ground disconnected, 0.65mA
Maximum permissible leakage, 3.5mA
Dimensions
Interior
21.3” W x 26.8” H x 20.0” F-B
(54.1cm x 68.1cm x 50.8cm)
Exterior
26.3” W x 39.5” H x 25.0” F-B
(66.8cm x 100.3cm x 63.5cm)


Water Jacketed CO
2
Incubator        7-3
Thermo Scientific
Section 7
Specifications
Weight (per unit)
Net
265 lb. (120.2 kg)
Net Operational
365 lb. (165.6 kg)
Shipping
324 lb. (147.0 kg)
Safety Specifications
Altitude
2000 meters
Temperature
5°C to 40°C
Humidity
80% RH at or below 31°C, decreasing linearly to 
50% RH at 40°C
Mains Supply Fluctuations Operating Voltage Range
Installation Category 2
1
Pollution Degree 2
2
Class of Equipment
1
Installation category (overvoltage category) defines the level of transient overvoltage
which the instrument is designed to withstand safely. It depends on the nature of the
electricity supply and its overvoltage protection means. For example, in CAT II which is
the category used for instruments in installations supplied from a supply comparable to
public mains such as hospital and research laboratories and most industrial laboratories,
the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V
supply.
2
Pollution Degree describes the amount of conductive pollution present in the operating
environment. Pollution Degree 2 assumes that normally only non-conductive pollution
such as dust occurs with the exception of occasional conductivity caused by condensation.


7-4
Water Jacketed CO2 Incubator
Thermo Scientific
Section 7
Specifications
Unit Attachment
Location
Caution Do not drill deeper than 1/2 inch.  
Exterior wrap is 18 ga. cold rolled steel. 

Single Units
Stacked Units


Water Jacketed CO2 Incubator        7-5
Thermo Scientific
Section 7
Specifications
Unit Center of Gravity
Locations
Single Units
Stacked Units


 


Section 8  
Spare Parts
Part #
Description
360171
Liquid Level Switch
191534
Feather Gasket
113002
5/16-18 Glide Foot
132046
115/230V Dual Heater
132049
Face Heater 40W, 24VAC
190730
Heated Inner Door
190619
Left Inner Door Hinge
190618
Right Inner Door Hinge
700013
0.500” Flanged Nylon Bearing
990026
Door Gasket w/ Magnet
290184
Temp Probe 2252 Ohm
191634
Micro Board
190609
Display Board
230153
6A Circuit Breaker/Switch
460157
Line Filter/Power Inlet
420096
130VA Transformer, Int’l, 14/28V S
230135
1 AMP Fuse for 115V Outlet
230159
3.5 Amp Fuse - Microboard
230158
2.5 Amp Fuse - Microboard
250087
Solenoid Valve Kit w/assorted fittings
1900071
Motor 2-Pole 115VAC
100113
Blower Wheel 3.5x1.5 CCW
190846
Blower Scroll Assembly
103072
Blower Plate Gasket
290090
CO
2
Sensor Assembly
103074
CO
2
Sensor Plate Gasket
130097
#6 Silicone Stopper w/ 3/8” Hole
180001
Polypropylene Funnel
430108
Line Cord Set
110084
Drain Plug
Water Jacketed CO
2
Incubator        8-1
Thermo Scientific
All Models


Spare Parts - Data Output
Options
Spare Parts - Humidity
Readout (190643)
Spare Parts - Gas Guard
(190640/190642)
Spare Parts - IR or O
2
(3120/3121, 3130/3131,
3140/3141)
Spare Parts - 230V 
(3111, 3121, 3131, 3141)
8-2
Water Jacketed CO
2
Incubator
Thermo Scientific
Section 8
Spare Parts
Part #
Description
770001
Bacterial Air Filter (CO
2
line, air, sample and access port)
760175
HEPA Filter
760199
HEPA2 VOC Filter Accessory Kit 
(760200 HEPA
2
,  117036 silicone plug, 101018 O-ring)
1900067
Filter Replacement Kit  
([2] 770001 filter and 760175 HEPA)
1900094
HEPA2 VOC Filter Replacement Kit 
(760200 HEPA2, 190985 access port filter, 770001 filter)
130096
Clean Stick (for water reservoirs, pans, and bottles)
130101
Clean Stick (quantity of 6)
190566
Inner Door Heater Cable
420097
43VA Transformer, INT. SRS
460138
Power Outlet, Snap-In Receptacle
230120
1/2 AMP Fuse for 230 V outlet
190885
IR Sensor
191646
O
2
Circuit Board
250119
AC Solenoid
770001
Filter
250118
Valve, Sol, O
2
, 12VDC, 4W
290083
O
2
Sensor Fuel Cell
190661
IR Sensor Filter
250121
Solenoid Valve, 3-way, 12VDC
360213
Pressure Switch
290154
RH Sensor Assembly, 1 ft.
190512
Analog Output Board 4-20 mA
190543
Analog Output Board 0-5V
190544
Analog Output Board 0-1V
73041
24 Ga. 3 Conductor, Analog Output Wire
190523
RS485 Output Kit
All Models (continued)


W
ater Jacketed CO
2
Incubator        8-3
Thermo Scientific
Section 8
Spare Parts


8-4
W
ater Jacketed CO2 Incubator
Thermo Scientific
Section 8
Spare Parts


Water Jacketed CO2 Incubator        8-5
Thermo Scientific
Section 8
Spare Parts


8-6
Water Jacketed CO2 Incubator
Thermo Scientific
Section 8
Spare Parts


W
ater Jacketed CO2 Incubator        8-7
Thermo Scientific
Section 8
Spare Parts


 


Water Jacketed CO
2
Incubator        9-1
Thermo Scientific
Section 9
Electrical Schematics


9-2
Water Jacketed CO
2
Incubator
Thermo Scientific
Section 9
Electrical Schematics


Water Jacketed CO
2
Incubator        9-3
Thermo Scientific
Section 9
Electrical Schematics


 


Water Jacketed CO2 Incubator        9-4
Thermo Scientific
Section 9
Electrical Schematics


9-5
Water Jacketed CO2 Incubator
Thermo Scientific
Section 9
Electrical Schematics


Water Jacketed CO2 Incubator        9-6
Thermo Scientific
Section 9
Electrical Schematics


 


Water Jacketed CO
2
Incubator        10-1
Thermo Scientific
THERMO 
FISHER SCIENTIFIC W
ATER JACKETED INCUBA
TOR W
ARRANTY USA*
The W
arranty Period starts two weeks from the date your equipment is shipped from our facility
. This allows shipping time so th
e warranty will go into
effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner duri
ng the first year war-
ranty period.
During the first year
, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at 
Thermo's expense,
including labor
. In addition, after the initial warranty period of one year
, the CO2 sensor (IR)  will be warranted for a secon
d year
, parts only
. The CO2
sensor (T/C) will be warranted for four additional years for a total of five years, parts only
. The water jacket chamber has a 
unit production lifetime war-
ranty
, parts only
, against rust-through and leaks. Unit production life is defined as the period of time the current chamber de
sign is in production, plus
seven years. If corrosive chemicals are identified in the chamber
, the warranty will be voided. Installation and calibration is
 not covered by this warranty
agreement. The T
echnical Services Department must be contacted for warranty determination and direction prior to performance of any repairs.
Expendable items, glass, filters and gaskets are excluded from this warranty
.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipmen
t or to the component
part beyond the original warranty period. The T
echnical Services Department must give prior approval for return of any componen
ts or equipment. At
Thermo's option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destina
tion.
THIS W
ARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER W
ARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED.  NO W
ARRANTIES
OF MERCHANT
ABILITY OR FITNESS FOR A P
ARTICULAR PURPOSE SHALL APPL
Y. 
Thermo shall not be liable for any indirect or consequential
damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. 
Printed instruction manu-
als carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your T
echnical Services Department at l-800-438-4851 (USA and Canada) or 1-740-37
3-4763. W
e're ready to
answer your questions on equipment warranty
, operation, maintenance, service and special applications. Outside the USA, contact
 your local distributor
for warranty information.
ISO 9001
REGISTERED
Rev
. 6  4/09


10-2
Water Jacketed CO
2
Incubator
Thermo Scientific
THERMO FISHER SCIENTIFIC INTERNA
TIONAL W
ATER JACKETED INCUBA
TOR W
ARRANTY*
The W
arranty Period starts two months from the date your equipment is shipped from our facility
. This allows shipping time so t
he warranty will go into
effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner duri
ng the first year warranty
period.
During the first year
, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at 
Thermo's expense, excepting
labor
. In addition, after the initial warranty period of one year
, the CO
2
sensor (IR) will be warranted for a second year
, parts only
. The CO
2
sensor (T/C) will
be warranted for four additional years for a total of five years, parts only
. The water jacket chamber has a unit production lifetime warranty
, parts only
,
against rust-through or leaks. Unit production life is defined as the period of time the current chamber design is in productio
n, plus seven years. If corrosive
chemicals are identified in the chamber
, the warranty will be voided. Installation and calibration is not covered by this warranty agreement. The T
echnical
Services Department must be contacted for warranty determination and direction prior to performance of any repairs. Expendable 
items, glass, filters and
gaskets are excluded from this warranty
.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipmen
t or to the component part
beyond the original warranty period. The T
echnical Services Department must give prior approval for return of any components or
 equipment. At Thermo's
option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destination.
THIS W
ARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER W
ARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO W
ARRANTIES OF
MERCHANT
ABILITY OR FITNESS FOR A P
ARTICULAR PURPOSE SHALL APPL
Y.
Thermo  shall  not be liable for any indirect or consequential dam-
ages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. 
Printed instruction manuals
carefully detail equipment installation, operation and preventive maintenance.
If equipment service  is required, please call your T
echnical Services Department at 1-800-438-4851 (USA and Canada) or 1-740-373-4763. W
e're ready to
answer your questions on equipment warranty
, operation, maintenance, service and special application.  Outside the USA, contact your local distributor for
warranty information.
Rev
. 6  4/09
ISO 9001
REGISTERED


Water Jacketed CO
2
Incubator        11-1
Thermo Scientific
H7
TURCK Inc.  3000 Campus Drive  Minneapolis, MN 55441   Application Support: 1-800-544-PROX   Fax: (763) 553-0708   www.turck.com
UNLOCK
LOCK
picofast “Snap Lock” Connector Instructions
Installation Instructions
To Attach:
1. Be sure black locking
sleeve is pulled back.
2. Line up pins and push
connector onto plug.
You will feel a “snap.”
Do not twist.
3. Push locking sleeve forward
on connector until it is
flush with the front of the
connector.
4.
If sleeve is difficult to slide
on a new connector,
“exercise” it a few times.
Do not use tools.
With TURCK’s patented Locking Sleeve pulled back,
any
picofast sensor slides on without any difficulty.
After sensor is connected, push Locking Sleeve for-
ward to create a watertight connection.   It’s a snap!
1. Do not twist.
2. Pull locking sleeve back
to “unlocked” position.
3.
Pull connector
straight off.
4.
Leave locking sleeve in
unlocked position.
To Detach:


 


Water Jacketed CO2 Incubator        12-1
Thermo Scientific


12-2
Water Jacketed CO2 Incubator
Thermo Scientific


Thermo 
Fisher Scientific
Controlled Environment Equipment
401 Millcreek Road, Box 649,
Marietta, Ohio 45750
United States
www.thermo.com

Document Outline

  • Table of Contents
  • Installation and Start-Up
  • Calibration
  • Configuration
  • Alarms
  • Preventive Maintenance
  • Routine Maintenance