Introduction
The radioactive actinide element neptunium (Np) is formed by neutron
bombardment of uranium (U), with more than 50,000 kg of Np produced
annually as a by-product of nuclear-power generation. Smaller quantities
of Np can be found as a decay product of americium-241 (
241
Am) used in
ionizing smoke detectors, with trace quantities being produced from nuclear
bomb testing, and naturally (from natural neutron capture) in U ores.
Accurate analysis of neptunium 237 in
a uranium matrix, using ICP-QQQ with
MS/MS
Application note
Authors
Garry Duckworth,
Springfields Fuels Ltd, UK,
Glenn Woods,
Agilent Technologies, UK,
Nuclear, environmental
2
The predominant isotope
237
Np is produced as follows:
237
Np sits at the top of the “neptunium decay series”
which in turn produces a series of other radioactive
elements with medium to very short half-lives,
eventually forming Bi and Tl (Figure 1).
237
Np is extremely mobile in the environment, as it
readily forms aqueous solutions (more so than any other
actinide), attaches to particles and colloids, and does
not readily become trapped in humic substrates such as
soil.
Np also has a high affinity for calcium-rich materials
including certain clays and concrete, so care is needed
with its storage. Although Np is not readily absorbed in
the human gut, once in the body, it will pre-concentrate
in the bones.
The relatively long half-life of
237
Np (~2.14 million
years) means that it is persistent in the environment,
so it requires suitable containment, and needs to be
monitored at low levels.
Abundance sensitivity of ICP-QQQ with MS/MS
Samples that contain Np usually contain U at far higher
concentrations. The determination of ultra-trace
237
Np
in environmental samples and nuclear materials (fuel or
waste) is difficult by ICP-MS because of the overlap due
to peak tailing from the large
238
U peak.
Conventional quadrupole ICP-MS instruments (ICP-
QMS) operate at unit mass resolution to separate
adjacent masses, meaning a nominal resolution
(expressed as M/ΔM) of 237 at m/z 237. However, peak
separation also depends on the abundance sensitivity
(AS) of the spectrometer. AS is a measure of the peak
tailing, calculated from the contribution that an intense
peak makes to its neighbouring masses.
Figure 1. Np-237 decay series. Adapted from “The NUBASE Evaluation of
Nuclear and Decay Properties”. Nuclear Physics A 729: 3–128. DOI:
10.1016/j.nuclphysa.2003.11.001
Quadrupole ICP-MS instruments can achieve AS of up
to 1 x 10
-7
(a peak of 1 x 10
7
cps contributes 1 cps to
the adjacent masses), so a peak with an intensity much
higher than 10
7
cps will make a significant contribution
to the peaks either side. High Resolution-Sector Field
(HR-SF)-ICP-MS has better resolution than ICP-QMS
(M/ΔM of up to 10,000), but poorer AS. So adjacent
peaks may appear to be separated on the mass scale,
but the peak tail of an intense peak may still contribute
to the masses above and below. The Agilent 8800 or
8900 Triple Quadrupole ICP-MS (ICP-QQQ) use a unique
configuration with two quadrupole mass analyzers (Q1
and Q2) either side of the collision reaction cell. When
both quadrupoles are operated as unit mass filters
(MS/MS mode), this configuration delivers unmatched
peak separation because the abundance sensitivity
performance is the product of two mass separations –
Q1 AS x Q2 AS – giving an overall AS of <<10
-10
.
3
ICP-QQQ is therefore able to successfully separate
237
Np
from the
238
U overlap, even when the U is present at
many orders of magnitude higher concentration. This
is demonstrated in Figure 2, which shows the spectra
of 100 ppt Np in a 10 ppm U matrix measured in Single
Quad (SQ) mode (top) and MS/MS mode below. The
ICP-QQQ spectrum shows the superior peak separation
provided by MS/MS mode, and the clear elimination
of the contribution on mass 237 from the adjacent 238
peak.
Figure 2. Spectra of 100 ppt Np in a 10 ppm U matrix sample solution
obtained using ICP-QQQ in Single Quad mode (top) and MS/MS mode
(bottom). MS/MS mode eliminates the peak tail on the low mass side of the
intense
238
U peak.
Experimental
Instrumentation
The Agilent 8800* ICP-QQQ was configured with an
SPS 4 autosampler, and the standard sample
introduction system consisting of a Micromist nebulizer,
quartz spray chamber, quartz torch and Ni interface
cones. Instrument operating parameters are given in
Table 1.
Table 1. ICP-QQQ operating parameters.
Parameter
Value
RF power
1550 W
Sampling depth
8.0 mm
Nebulizer gas flow rate
1.15 L/min
Spray chamber temp
2 °C
Cell gas
None
Calibration
A blank and four neptunium calibration standards
from 100 to 2000 ng/L (ppt) were prepared in nitric acid
(2% v/v, Ultrapur, Merck, Germany ). The calibration
curve obtained by ICP-QQQ showed excellent linearity
over the calibrated range with a calibration coefficient
of 1.0000 (Figure 3). As Np is essentially absent from
the environment and does not occur as a typical reagent
contaminant, the Background Equivalent Concentration
(BEC) obtained was ~0.0009 ng/L (0.9 pg/L, ppq) and
the Detection Limit (DL) was ~0.0031 ng/L (3.1 ppq).
This illustrates the exceptionally low background and
high ion transmission (sensitivity) of the ICP-QQQ when
operating in MS/MS mode.
Figure 3. Calibration curve for
237
Np obtained by ICP-QQQ.
235
U
234
U
238
U
237
Np cannot be measured
due to overlap from the
238
U peak
237
Np
* The Agilent 8800 ICP-QQQ has been superseded by the 8900 ICP-QQQ
4
Results and discussion
Table 2. Measurement of
237
Np in a series of U matrix samples using ICP-QQQ and ICP-QMS.
8800 ICP-QQQ
7900 ICP-QMS
Sample Name
Reported
237
Np
conc., ug/L
CPS
Reported
237
Np
conc., ug/L
CPS
1 ppm U - unspiked
0.0000__0.83__0.0154__5519.75'>0.0000
1.10
0.0016
570.48
1 ppm U, 0.1 ppb Np
0.1021
14942.85
0.1018
36525.19
1 ppm U, 1.0 ppb Np
1.0445
152806.38
1.0100
362304.66
10 ppm U - unspiked
0.0000
0.83
0.0154
5519.75
10 ppm U, 0.1 ppb Np
0.1029
15052.99
0.1152
41339.80
10 ppm U, 1.0 ppb Np
1.0486
153402.27
1.0196
365764.86
100 ppm U - unspiked
0.0000
3.97
0.1581
56728.76
100 ppm U, 0.1 ppb Np
0.0997
14586.02
0.2494
89482.09
100 ppm U, 1.0 ppb Np
0.9859
144228.95
1.0597
380137.27
The Abundance Sensitivity performance of the ICP-QQQ
in MS/MS mode was tested using a series of spiked
and unspiked uranium solutions. Three sets of uranium
solutions were prepared, at concentrations of 1, 10
and 100 mg/L (ppm). For each concentration level, the
U matrix solutions were measured unspiked, and with
Np spikes at 0.1 and 1.0 µg/L (ppb). For comparison
purposes, the samples were analyzed using an Agilent
8800 ICP-QQQ and an Agilent 7900 ICP-QMS, to assess
the impact of the improved AS performance of the
QQQ configuration.
The results in Table 2 show that accurate recoveries
were achieved for both Np spike levels in all of the U
matrix samples analyzed by ICP-QQQ—even with U:Np
concentrations at a ratio of 1,000,000:1. In contrast, the
Np results measured on the 7900 quadrupole ICP-MS
show a small contribution from the U matrix. This U
signal contributes to a false-positive result for Np in
the higher U matrix samples, including the unspiked
U matrices. While the U signal only increased the
apparent Np concentration by a small amount
(sub-µg/L), the low level at which Np must be
monitored means that this false signal is significant.
The ICP-QMS results show that the interference
effect is more pronounced with increasing U matrix
concentration, due to the relatively poor AS of ICP-QMS
compared to ICP-QQQ.
Conclusions
ICP-MS is used successfully for the analysis of trace
elements in a wide range of complex sample matrices.
However, several challenging interferences remain,
including the measurement of trace analytes that occur
close to major or matrix element peaks. This study has
shown that the superior abundance sensitivity provided
by the ICP-QQQ’s tandem quadrupole mass analyzer
configuration (MS/MS) has practical benefits for the
analysis of trace concentrations of Np at m/z 237 in the
presence of high concentrations of the adjacent major
isotope of U at m/z 238.
Conventional quadrupole-ICP-MS cannot resolve the
overlap/peak tailing from the
238
U isotope sufficiently to
allow ultra-trace level analysis of
237
Np; SF-ICP-MS also
has insufficient abundance sensitivity to resolve the
adjacent masses well enough to perform this analysis.
www.agilent.com/chem
Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance or use of this material.
Information, descriptions, and specifications in this publication are subject to change
without notice.
© Agilent Technologies, Inc. 2017
Published December 19 2017
Publication number: 5991-6905EN
Dostları ilə paylaş: |